Centralized control, easy to manage.
The quality of dialysate can be effectively improved by adding precision filter in the supply line.
Monitoring Advantage.
It is convenient to monitor the ion concentration of dialysate and avoid the single machine dispensing error.
Centralized Disinfection Advantage.
After dialysis every day, the system can be disinfected in linkage without blind spots. The effective concentration and residual concentration of disinfectant are easy to detect.
Eliminate the possibility of secondary pollution of concentrate.
Current use after mixing, reducing biological pollution.
Save cost: Reduced transportation, packaging, labor costs, reduced space for concentrate storage.
Product Standard
1. The overall design conforms to the health standard.
2. Product design materials meet the requirements of hygiene and corrosion resistance.
3. Preparation of concentrate: water inlet error ≤ 1%.
Safety Design
Nitrogen generator, effectively inhibit the growth of bacteria.
Liquid A and liquid B operate independently, and they are composed of liquid distribution part and storage and transportation part respectively. Liquid distribution and supply do not interfere with each other and will not cause cross contamination.
Multiple safety protection: ion concentration monitoring, endotoxin filter and pressure stabilizing control to ensure the safety of patients and dialysis equipment.
Eddy current rotary mixing can fully dissolve powder A and B. Regular mixing procedure and prevent the loss of bicarbonate caused by excessive mixing of B solution.
Filter: filter the undissolved particles in the dialysate to make the dialysate meet the requirements of hemodialysis and effectively ensure the quality of the concentrate.
Full circulation pipeline is used for liquid supply, and circulation pump device is installed to ensure the stability of liquid supply pressure.
All valves are made of anti-corrosion materials, which can withstand long-term immersion of strong corrosive liquid and have long service life.
Automatic Control
After dialysis every day, the system can be disinfected in linkage. There is no blind spot in disinfection. The effective concentration and residual concentration of disinfectant are easy to detect.
Fully automatic liquid preparation program: working modes of water injection, timing mixing, filling the liquid storage tank etc., to reduce the use risk caused by insufficient training.
Fully automatic washing and one key disinfection procedures to effectively prevent bacterial growth.
Personalized Installation Design
A and B liquid pipelines can be laid according to the requirements of the actual site of the hospital, and the pipeline design adopts the full cycle design.
Liquid preparation and storage capacity can be selected at will to meet the needs of departments.
Compact and integrated design to meet the combined installation requirements of various site conditions.
Power supply | AC220V±10% |
Frequency | 50Hz±2% |
Power | 6KW |
Water requirement | temperature 10℃~30℃, water quality meets or better than the requirements of YY0572-2015 "water for Hemodialysis and Relate Treatment. |
Environment | Ambient temperature is 5℃~40℃, relative humidity is not greater than 80%, atmospheric pressure is 700 hPa~1060 hPa, no volatile gas such as strong acid and alkali, no dust and electromagnetic interference, avoid direct sunlight, and ensure good air mobility. |
Drainage | drainage outlet ≥1.5 inches, the ground needs to do a good job of waterproof and floor drain. |
Installation: installation area and weight | ≥8(width x length =2x4) square meters, the total weight of the equipment loaded with liquid is about 1 ton. |
1. Preparation of concentrated liquid: automatic water inlet, water inlet error ≤1%;
2. The preparation solution A and B is independent of each other, and consists of liquid mixing tank and storage with transportation respective. The mixing and supplying parts do not interfere each other;
3. The preparation of concentrated solution is fully controlled by PLC, with 10.1 inch full-color touch screen and simple operation interface, which is convenient for medical staff operating;
4. Automatic mixing procedure, working modes such as water injection, timing mixing, perfusion; Fully dissolve A and B powder, and prevent the loss of bicarbonate caused by excessive stirring of B liquid;
5. Filter: filter the undissolved particles in the dialysis solution, make the dialysis solution meet the requirement of hemodialysis, effective ensure the quality of concentrated solution;
6. Fully automatic flushing and one-button disinfection procedures, effectively prevent the breeding of bacteria;
7. Opened disinfectant, the residual of the concentration after the disinfection meet the standard requirements;
8. All valve parts are made of corrosion-resistant materials, which can be soaked for a long time by strong corrosive liquid and have long service life;
9. The product materials meets the requirements of medical and corrosion resistance;
10. Multiple safety protection: ion concentration monitoring, endotoxin filter, stable pressure control, to ensure the safety of patients and dialysis equipment;
11. Mixing according to the actual needing, reduce errors and pollution.